In the last part of our series of articles about trace metal analysis using solid-state electrodes, we will have a look at the glassy carbon rotating disc electrode (GC RDE) and its application possibilities.

Did you read the other parts in this series? Find them here!

The Glassy Carbon Rotating Disc Electrode

A rotating disc electrode (RDE) consists of two parts: the electrode tip which is made available in different materials, and a driving axle. The electrode tip is simply screwed onto the axle (Figure 1) to assemble the complete working electrode.

Figure 1. The two parts which make up the RDE. Left: driving axle for RDE. Right: glassy carbon electrode tip, with shaft made of glass.

Glassy carbon (GC) has a long history as solid electrode material for trace metal analysis. In general, GC is carbon with an amorphous structure which is similar to glass or ceramics, but different from graphite or diamond which both have a crystalline structure.

Aside from properties including a high temperature stability and a hardness similar to quartz, glassy carbon is very chemically inert and has a low electrical resistance, making it a versatile electrode material.

In the Metrohm GC electrode tip (Figure 1), the glassy carbon rod is fused within a glass shaft—another inert material. This design creates an electrode tip that is inert against most chemicals and solvents and guarantees measurements with excellent reproducibility due to the seamless intersection between the electrode material and glass shaft.

Modification with a metal film

For trace metal applications, the GC electrode is modified with a metal film, usually mercury or bismuth. The film is plated ex-situ from an acid plating solution which contains about 20 mg/L Hg2+ or Bi3+. Such a solution can easily be prepared from commercially available metal standard solutions and can be used for the plating of several films.

Once the film is deposited on the glassy carbon electrode, multiple determinations can be carried out with the same film. When the performance deteriorates, the exhausted film is simply wiped off and a fresh film is plated. Since only the renewable film is affected by aging processes, the GC electrode itself can be used for a very long time.

Applications using glassy carbon electrodes exhibit excellent reproducibility and stability in combination with very low detection limits.

Figure 2. Glassy carbon rotating disc electrode in a 884 Professional VA instrument from Metrohm.

Applications

Cadmium and lead determinations

The risk of cadmium and lead poisoning from drinking water and the significance of the determination of these two elements has already been discussed in previous posts in this series. To monitor the guideline values of 3 µg/L for cadmium and 10 µg/L for lead, recommended by the WHO (World Health Organization), a detection limit of β(Cd) = 0.3 µg/L and β(Pb) = 1 µg/L would be sufficient.

With the glassy carbon electrode the determination is far more sensitive, featuring a ten-fold improvement on the limit of detection of β(Cd) = 0.02 µg/L and β(Pb) = 0.05 µg/L with a deposition time of 30 s. This limit can be lowered even more with an increased deposition time.

For this extremely sensitive determination, a mercury film is plated on the glassy carbon electrode. The determination of cadmium and lead is carried out by anodic stripping voltammetry (ASV).

To learn more about this application, please check our website.

Free Application Note download: AN-V-225 Cadmium and lead in drinking water – Simultaneous determination on a mercury film modified glassy carbon electrode.

The very low detection limit makes this application especially interesting when it is not only required to monitor limit values but to actually detect concentrations in the ppt (parts per trillion, ng/L) range, e.g. in environmental analysis such as for seawater research.

Nickel and cobalt measurements

Another application with very low detection limits using the GC electrode is the determination of nickel and cobalt. This electrode allows the detection of concentrations down to β(Ni) = 0.05 µg/L and β(Co) = 0.03 µg/L. For this application, the electrode is modified with a bismuth film. The determination of nickel and cobalt is carried out by adsorptive stripping voltammetry (AdSV) using the complexing agent DMG (dimethylglyoxime).

Figure 3. Determination of β(Ni) = 0.34 µg/L and β(Co) < LOD in tap water (30 s deposition time) using the GC RDE.

For decades, this method was successfully executed with the mercury drop electrode. The use of a bismuth film on a glassy carbon electrode offers a non-toxic alternative with a similar sensitivity as the established method. Besides the high sensitivity, this application also shows excellent repeatability.

20 consecutive determinations of β(Ni) = 0.5 µg/L and β(Co) = 0.5 µg/L, carried out on the same bismuth film, showed an average recovery of 105% for nickel, with a relative standard deviation (RSD) of 2.0%. The recovery for cobalt was 112% with a RSD of 3.3%. This makes this method a viable tool in environmental analysis when natural background concentrations, which are often in the ppt (ng/L) range, should be investigated.

For further details about this application, please refer to Application Note AN-V-224: Nickel and cobalt in drinking water – Simultaneous determination in low ng/L range on the GC RDE modified with a bismuth film.

Chromium(VI) monitoring

Legal limits for chromium are relatively high. For example, the guideline value of the World Health Organization (WHO) is 50 µg/L for drinking water. These values usually refer to the total chromium concentration, but there are significant differences in toxicity between Cr(III) and Cr(VI). Even miniscule doses of Cr(VI) are toxic as well as carcinogenic.

Since the beginning of this century, there have been ongoing discussions in the scientific community about whether an additional limit value only for Cr(VI) is required, and what this value should be.

Measuring techniques are needed which allow the determination of Cr(VI) in the ng/L range. Using the glassy carbon electrode modified with a mercury film it is possible to detect Cr(VI) concentrations down to 0.05 µg/L. Cr(VI) is determined by adsorptive stripping voltammetry (AdSV) with DTPA (diethylenetriaminepentaacetic acid) as complexing agent. The recovery of a concentration of β(Cr(VI)) = 0.1 µg/L is 111% with a relative standard deviation of 4.4% (triplicate determination).

If you are interested to learn more, download our free Application Note V-277: Chromium(VI) in drinking water – Ultra-sensitive determination on the mercury film modified glassy carbon electrode (DTPA method).

All the above-mentioned applications can be carried out manually with a 884 Professional VA system (Figure 4), but it is also possible to run small sample series with an automated setup.

Figure 4. 884 Professional VA with two 800 Dosinos for automatic addition of electrolyte and standard solution.

Summary

This was the last post in our five-part series on heavy metal analysis with solid state electrodes. If this or one of the previous posts sparked your interest in one of the applications, do not hesitate to contact your local Metrohm representative.

For a complete overview of the different applications that can be performed with the SSEs exhibited in this series, check out the table below. Click on each application note or bulletin for a free download! 

Overview: Applications with Metrohm SSEs
Element Electrode Application Document Lab Portable
Ag GC RDE Application Bulletin 207

As scTRACE Gold Application Note V-210
Application Note V-211

Bi scTRACE Gold Application Note V-218

Cd, Pb GC RDE (Hg film) Application Note V-225

Cd, Pb SPE (Hg film) Application Note V-231

Cd, Pb Bi drop Application Note V-221

Cr(VI) GC RDE (Hg film) Application Note V-227

Cr(VI) scTRACE Gold (Hg film) Application Note V-230

Cu scTRACE Gold Application Note V-213

Fe scTRACE Gold Application Note V-216

Fe Bi drop Application Note V-222

Hg scTRACE Gold Application Note V-212

Ni, Co scTRACE Gold (Bi film) Application Note V-217

Ni, Co GC RDE (Bi film) Application Note V-224

Ni, Co SPE (Bi film) Application Note V-232

Ni, Co Bi drop Application Note V-223

Pb scTRACE Gold (Ag film) Application Note V-214

Sb(III) scTRACE Gold Application Note V-229

Se(IV) scTRACE Gold Application Note V-233

Te(IV) scTRACE Gold Application Note V-234

Tl scTRACE Gold (Ag film) Application Note V-228

Zn scTRACE Gold Application Note V-215

Post written by Barbara ZumbrägelProduct Manager VA/CVS at Metrohm International Headquarters, Herisau, Switzerland.

%d bloggers like this: