Select Page
Best practice for separation columns in ion chromatography (IC) – Part 3

Best practice for separation columns in ion chromatography (IC) – Part 3

This is the third and final post in our series providing you with tips and tricks on the proper use of ion chromatography columns. In the first part, we mostly discussed the standard operation conditions as well as operational limits for columns, while in the second post, we focused on application related topics and what elution parameters can be changed to modify the separation performance. In the conclusion of this series we will take a closer look at the ways to assess the column performance during its lifetime and offer some troubleshooting tips which can help in fixing issues that may appear. 

Essential column parameters

Let’s begin by looking at the main parameters that can be used to judge the column performance. Most of these reference values can be extracted from the certificate of analysis. The separation column certificates can be found in the Metrohm Certificate Finder. Reproducing the certificate chromatogram periodically to verify the column performance can be helpful to detect changes in performance early and to avoid irreversible damage of the column.

Below, we will consider the most important parameters one by one.
Retention time and column capacity

The retention time of the individual analytes is a good measure of the column selectivity and capacity. When the retention times of the analytes do not match the retention times of the certificate, there are many possible reasons for this behavior.

Possibly, the eluent composition is wrong, e.g., because the eluent components are not present in the correct concentrations. In case the eluent strength is too high, the retention times will be reduced. This phenomenon usually occurs for all ions simultaneously and shifts all the peaks closer together. Thereby, the multivalent ions are accelerated to a stronger extent than the monovalent ions. This issue can be fixed by preparing a fresh eluent with the correct composition.

The eluent may not be fresh or not sufficiently protected from the surrounding atmosphere, e.g., with a CO2 adsorber. Carbon dioxide from the ambient air can change the eluent composition (i.e., strength, pH, etc.) over time and this will affect the retention times of the analytes. Weak hydroxide eluents are particularly affected by this issue, since the elution strength of carbonate ions is much stronger than the one of hydroxide ions, leading to strong shifts in retention times. Therefore, multivalent ions are more affected by this than monovalent ones. This issue can be fixed by preparing a fresh eluent and by using a CO2 adsorber. When working with hydroxide eluents, it is also important to verify the status of the hydroxide stock solution as it can absorb CO2 from the air. Figure 1 shows the effect of CO2 uptake from the air on the retention times of common anions.

Effect of eluent absorption of carbon dioxide in the air on ion separation.
Figure 1. Example chromatograms showing the effect of eluent absorption of carbon dioxide in the air for seven standard anions measured on a Metrosep A Supp 5 – 250/4.0 column under standard conditions. Red: unprotected eluent. Black: eluent protected with a carbon dioxide adsorber.
Some samples may contain constituents that occupy the ion exchange groups of the column material. In that case, the number of available ion exchange groups on the column is reduced compared to the initial state, leading to an apparent smaller column capacity and shorter retention times. Depending on the type of contaminant(s), there are possibilities to reverse this effect and wash them from the column surface. These regeneration procedures are adapted individually to the different stationary phases and should be used as a last option to save a column.
The procedures as well as the operation limits are described in the respective leaflets of the column and can be found on our website. Please consult your Metrohm dealer before performing a regeneration procedure.
  • In case the contamination is from a multivalent ion, it is usually possible to wash them off by flushing the column with an eluent of a higher concentration. When doing so, do not forget the operation limits of the column, particularly regarding the eluent pH.
  • Sometimes, organic molecules can adsorb to the column stationary phase because of their strong affinity to it. This can lead to a blockage of the ion exchange groups and reduced retention times along with increased operating pressure. When this happens, it can help to wash the stationary phase with an eluent containing a fraction of an organic modifier. This will increase the affinity of the contaminant to the mobile phase and help to detach it from the stationary phase.

Usually, after such a regeneration procedure, the original column capacity can be restored.

An even better way to extend the lifetime of the separation column is by regularly exchanging the guard column. One of its purposes is to protect the separation column from contaminants by retaining them.

Make sure to exchange the guard column three to four times over the lifetime of the separation column, or even more frequently when dealing with very complex sample matrices such as dyes or food matrices. In these situations, we suggest using the Metrosep RP 2 Guard/3.5.

The chromatogram overlay in Figure 2 shows the effect of a defective guard column on the peak shape.
Effect of eluent absorption of carbon dioxide in the air on ion separation.
Figure 2. Example chromatograms showing the effect of a damaged guard column. Black: analysis of standard anions when using a defective guard column. Red: identical analysis without a guard column installed. In this case, the replacement of the guard column is strongly recommended.
In some cases, the column capacity is reduced due to chemical modification of the stationary phase. This happens when a column is used in conditions outside of the operation limits—e.g., extreme pH situations. Extreme pH conditions can alter the chemical bonds of the ion exchange groups to the base particle substrate. This irreversible process leads to permanent column capacity loss. In this case, no regeneration procedure can restore the original column capacity.

Please note that this loss of ion exchange groups also happens slowly during the regular use of ion exchange columns. The strength of the chemical bond of the ion exchange groups is thoroughly tested during the development of Metrohm columns to provide a long column lifetime.

Theoretical plates and peak symmetry

The number of theoretical plates (TP) can be a useful tool to judge the column packing status. The higher the TP value, the better the packing of the column. For a meaningful judgement of the column packing bed, choose an analyte that does not elute too early in the chromatogram (these peaks can be affected by extra-column effects, in particular on 2 mm systems) and that is not affected by secondary interactions (i.e., avoid using nitrate). For cations, potassium is a good choice, while for anions, sulfate is a suitable choice.

Although the columns are packed at pressures higher than the normal operating pressure, the packing bed continues to further densify throughout the column lifetime with the continuous application of flowing mobile phase (eluent). While the packing bed of the stationary phase improves due to this effect, it may also lead to some dead volume at the entrance of the separation column. This dead volume can be responsible for peak broadening and reduced theoretical plates, particularly at the beginning of the chromatogram.

While this is a regular aging process of the column, it is possible to slow it down by taking good care of the column. Preventive actions include the slow initiation of the eluent flow rate and temperature at start-up as well as the proper and complete shutdown of the high-pressure pump before removing the column, as mechanical stress can be detrimental to the packing bed. Exchanging the guard column can often have a positive impact on the theoretical plates and peak shapes (see Figure 2).

Issues with the column packing bed often also become visible in the asymmetry factor values and the overall peak shapes. More pronounced fronting as well as peak broadening can be warning signs for channeling in the column or the guard column. Unfortunately, this kind of damage on the column is irreversible and requires an exchange of the separation column.

Before column replacement, it is important to check if the issues originated from the column or another connection in the IC system whenever peak broadening is observed in the chromatogram. Make sure that all capillaries in the high-pressure path have a diameter of ≤0.25 mm and that all capillaries have been installed and connected correctly without additional dead volume. Systems equipped with small inner diameter (2 mm, microbore) columns are more strongly affected by dead volume than those with 4 mm columns. This means that when using microbore columns, less dead volume is required for the peak broadening effect to become visible compared to when using 4 mm columns.

Column pressure

Another important parameter to regularly check throughout the column lifetime is the system pressure. High pressure is among the most frequent reasons leading to a column replacement. Whenever a pressure increase is noted, it is important to verify what part of the IC system is the cause.

If the sample contains particles and insufficient sample preparation is applied, the particles will accumulate at the entrance of the protective guard column and eventually lead to increased system pressure. In this case, the guard column performs its intended use of protecting the separation column and needs to be replaced. If the guard column is not replaced soon enough, then particulate contamination may break through and load onto the separation column. While there is no regeneration procedure for a guard column loaded with particles, separation columns contaminated with particles may be regenerated by rinsing the column at low flow rates in the reverse flow direction.

Please note that these regeneration procedures may not always be successful.

Avoiding particles from being injected into the high-pressure path of the IC system is a good way of protecting the guard and separation columns. Metrohm provides several automated sample preparation techniques which result in a positive impact on the column lifetime. The most common techniques for this purpose are Inline Ultrafiltration (Figure 3) and Inline Dialysis.

Effect of eluent absorption of carbon dioxide in the air on ion separation.
Figure 3. Inline Ultrafiltration is a useful automated sample preparation technique that protects the separation column from particle accumulation.
Learn more about Metrohm Inline Ultrafiltration in our related blog post.

Column end of life

All of the parameters described above should be considered to judge the column (and the IC system) performance. Many of these parameters can be monitored closely in the MagIC Net software so that a potential issue can be detected as early as possible.

Even though Metrohm IC columns are designed and manufactured to have very long lifetimes, at some point performance will decrease and even regeneration procedures may not be able to restore the required column performance to solve the application needs. This represents the end of the column life and makes the exchange of the separation column unavoidable.

Metrohm separation columns cannot be recycled and can be disposed with normal waste. However, depending on the samples measured as well as the used chemical types and their associated hazards, it may be necessary to consider a proper disposal option.

Irrespective of the sample nature, do not open the column at any time. 

Troubleshooting overview

Table 1 gives an overview about certain troubleshooting strategies when looking at your column performance behavior.

Table 1. Preventing and correcting performance loss in IC columns

 

Indicator Cause Preventive and corrective measures
Increasing counterpressure Particles on the guard column Replace the guard column.
Particles on the separation column

Rinse out the separation column in the reverse flow direction

  • Place the column inlet (i.e., the opening next to the intelligent chip) in a beaker, as this is now the outlet path for the contaminants.
  • Rinse out the separation column for approx. one hour.
  • Reinstall the separation column in the flow direction.
Particles in the sample Sample preparation, e.g., remove particles through lnline Ultrafltration
Shortened retention time Carbonate in the eluent

Carbon dioxide from the air affects the carbonate/hydrogen carbonate balance in the eluent. A carbonate/hydrogen carbonate eluent weakens over time; a hydroxide eluent strengthens.

  • Always tightly seal eluent bottles and bottles containing eluent concentrate.
  • Always use a CO2 adsorber.
Air bubbles in the eluent

Air bubbles make the eluent flow unstable. The counterpressure is an indicator of unstable flow. lt should remain stable
within a range of ± 0.1 MPa.

 

  • Deaerate the high-pressure pump.
  • Use an eluent degasser.
Capacity loss in the column due to high- valency ions Regenerate the column as per the column leaflet to remove any inorganic deposits.
Resolution loss Eluent too old or produced incorrectly Eluents should be freshly prepared. Make sure that they are produced correctly and particularly that carbonate and hydrogen carbonate are not confused.
Adsorptive effect of the contamination deposited in the guard column Replace the guard column.
Adsorptive effect of the contamination deposited in the separation column Regenerate the column as per the column leaflet to remove any organic or inorganic deposits.
Loss of theoretical plates Guard column contaminated Replace the guard column.
Separation column contaminated Regenerate the column as per the column leaflet to remove any organic or inorganic deposits.
Separation column overloaded

The separation column can be overloaded by factors such as a high salt content in the sample matrix.

  • Dilute the sample.
  • Inject less sample.
Dead volume in the IC system
  • Check that all capillaries have a diameter ≤0.25 mm; if they don’t, replace the capillaries.
  • Check that all of the capillaries have been installed correctly. The installation process is described step by step in the «IC Maintenance» multimedia guide.
Asymmetry Dead volume or contamination on the guard column Replace the guard column.
Separation column contaminated Regenerate the column as per the column leaflet to remove any organic or inorganic deposits.

 

Conclusion

This article explained how IC column performance can be assessed and monitored throughout the column lifetime and what measures can be taken to ensure a long column lifetime. With that, we now conclude the series «Best practice for separation columns in ion chromatography». If you have more questions, do not hesitate to contact your local Metrohm IC salesperson.

Get free tips and tricks

for your IC columns 
Post written by Dr. Vincent Diederich (Product Manager IC Columns) and Dr. Anne Katharina Riess (Head of Column Division) at Metrohm International Headquarters, Herisau, Switzerland.
Best practice for separation columns in ion chromatography (IC) – Part 3

Best practice for separation columns in ion chromatography (IC) – Part 2

The second part of this blog series about best practice for IC separation columns focuses on application related topics that have an impact on the column suitability and stability. First, there is the proper choice of the column that best suits the intended application. Then we turn to the operating parameters which can be modified in order to optimize the separation between analytes, and what the respective effects and possibilities are.

Choice of column length and diameter

Metrohm offers a broad range of columns that contain different stationary phases, have different lengths and/or inner diameters. The choice of the stationary phase has a great impact on the selectivity between the individual analytes on the one hand, as well as the stability against different sample matrices on the other hand. Instead, the column length has no impact on the selectivity, but rather on the separation efficiency between the individual peaks.

Find out more about Metrohm’s wide selection of separation columns for ion chromatography in our Column Catalog.
Effects of column length

In the following chromatograms (Figure 1), the effect of the column length on the separation efficiency for the Metrosep A Supp 17 column series is shown. Whenever choosing a column length, you should take the complexity of the intended separation and the presence of matrix components that could disturb the ions of interest into account.

Figure 1. Effect of column length on the retention times of the standard anions on the Metrosep A Supp 17 column (1: fluoride, 2: chloride, 3: nitrite, 4: bromide, 5: nitrate, 6: sulfate, 7: phosphate). Click image to enlarge.
Effects of column diameter

In addition to providing different lengths of IC separation columns, Metrohm also offers most columns in both in 4 mm inner diameter and 2 mm inner diameter (known as «microbore») versions. With regard to this, there are several criteria to distinguish:

  • If you use online systems in a continuous mode (i.e. systems which run unattended for several days in a row such as the Metrohm Process Analytics MARGA system – Monitor for AeRosols and Gases in Ambient air), we recommend using 2 mm IC columns. Due to the reduced flowrate for microbore columns (only 25% of the flowrate for 4 mm columns), the eluent and the regenerant solutions last much longer, which increases the time the instrument can be left unattended.
  • There are applications that require hyphenated techniques such as IC-MS for higher analyte selectivity and sensitivity. In this case, the use of 2 mm columns is ideal. The low flowrate is optimal for the electrospray process, and thus no flow splitter is required before entering the mass spectrometer.
  • Sometimes, only a limited amount of sample is available for injection. In these situations, 2 mm columns are preferred. This is because less dilution/diffusion occurs during the separation process and therefore higher signals are obtained.
  • On the other hand, if your sample contains a high load of matrix components, then selecting a suitable 4 mm IC columns will be a better choice because of the higher capacity available to separate the desired analytes from the matrix.
Find out more about MARGA and its capabilities for continuous air quality monitoring in our blog post.

Optimizing the analyte separation

Next to the column itself, several other parameters can be modified to optimize the selectivity of the separation. These parameters include temperature, eluent components and strength, and organic modifiers.

Effects of modifying the temperature

One of the simplest ways to fine tune the separation selectivity in IC is by modifying the temperature of the analysis. This is accomplished by using the integrated column oven in the instrument (if available). Multiple effects can be observed, for instance in anion analysis. As an example, the impact of the temperature on the selectivity is shown in the chromatogram overlay (Figure 2) for the Metrosep A Supp 17 column line.

Figure 2. Effect of temperature variation on the retention times of a suite of standard anions on the Metrosep A Supp 17 column (1: fluoride, 2: chloride, 3: nitrite, 4: bromide, 5: nitrate, 6: sulfate, 7: phosphate). Click image to enlarge.
  • The monovalent ions such as fluoride, chloride, nitrite, bromide, and nitrate are all accelerated with increasing temperature, indicating that fewer interactions with the stationary phase happen.
  • The behavior of multivalent ions such as phosphate or sulfate is more complicated to describe and will vary with each stationary phase. In general, multivalent ions are retarded more at higher temperatures, which causes the retention times to increase, as can be seen for sulfate. Phosphate on the other hand behaves differently, because of the temperature induced change of the eluent pH in a range close to the pKa value of phosphate. Due to this pH change, the effective charge of the phosphate ion changes as well (in this example, the effective charge is reduced with increasing temperature).
  • The peak shape of the polarizable ions such as nitrite, bromide, and in particular nitrate, is significantly improved at higher temperatures. The reason for this behavior is the reduction of secondary interactions with the stationary phase.
Effects of modifying the eluent composition and strength

Eluent composition and strength can be used to change the elution order of several analytes while using the same separation column. In cation chromatography, a retention model was developed by P.R. Haddad and P.E. Jackson, which allows researchers to predict retention times when changing the eluent composition [1].

Considering that the column remains identical in each determination, no change of ion exchange equilibrium and column capacity is to be expected. Therefore, when changing only the eluent concentration, the following correlation can be used:
Where:

  • k’ is the retention factor of the analyte of interest
  • c is a constant
  • x is the charge of the analyte
  • y is the charge of the eluent
  • Ey+M is the concentration of the eluent in the mobile phase
If nitric acid is used as the eluent, y = 1, and the model can be simplified to:
Applying this formula to practical situations in the laboratory means the following: with increasing the eluent strength, alkaline earth metals are accelerated much faster (x = 2) in comparison with alkali metals (x = 1), and thus it is possible to elute magnesium before potassium. This effect is called electroselectivity.

Multivalent metal ions are capable of forming complexes with dedicated complexing agents. Therefore, selectivities can be modified by adding complexing agents to the eluent. As an example, dipicolinic acid (DPA) is often used to complex calcium, which leads to a reduction of the effective charge of calcium. As a consequence, the retention time of calcium is reduced and calcium elutes before magnesium in the chromatogram (Figure 3).

The retention of monovalent cations can be influenced by the addition of crown ether to the mobile phase.

Figure 3. Effect of DPA concentration in the eluent on the retention times of several cations measured using the Metrosep C 6 column.
Anion systems are more complex regarding the retention time model, although the same electroselectivity effect can be observed to some extent for anions. However, when changing the eluent strength, the eluent pH also frequently changes, leading to different deprotonation equilibria of multivalent anions (e.g. phosphate). This influences the effective charge of the analyte, and by doing so, also influences its retention in a similar way as previously described for the effects of changing temperature.

In some cases, the use of a small amount of an organic modifier such as methanol, acetonitrile, or acetone in the eluent can make sense:

  • If bacterial contamination has been an issue before, the addition of 5% methanol to the eluent can help prevent future bacterial growth.
  • When samples containing a lot of organic solvent(s) need to be injected and no sample pretreatment such as extraction or matrix elimination (MiPCT-ME) is possible, it is recommended to add a suitable organic modifier to the eluent to ensure that the organic solvent(s) can be properly flushed out of the chromatographic column.
  • When using IC-MS, it is also recommended to add an organic modifier to the eluent to improve the electrospray process.

Be aware that the addition of organic modifiers will also affect the separation selectivities. For the standard anions, the effect is similar to that observed with increased temperatures: the peak shapes of the polarizable ions such as nitrite, bromide, and nitrate are improved.

Organic acids on the other hand may react very differently compared to the standard anions, and their reaction also strongly depends on the type of organic modifier used. Sample chromatograms that show the effect of the organic modifier on retention of analytes are shown in the manual for the Metrosep A Supp 10 column.

Download the Metrosep A Supp 10 column manual here to see example chromatograms showing the effects of organic modifier on analyte retention time.
For more information about column care, check out our blog post for different tips and tricks.

The History of Metrohm IC

Metrohm ion chromatography: bringing top quality and exceptional analytical performance to the lab since 1987. 
Reference

[1Haddad, P. R.; Jackson, P. E. Ion Chromatography: Principles and Applications; Journal of chromatography library; Elsevier; Distributors for the U.S. and Canada, Elsevier Science Pub. Co: Amsterdam, Netherlands; New York: New York, NY, USA, 1990.

Post written by Dr. Vincent Diederich (Product Manager IC Columns) and Dr. Anne Katharina Riess (Head of Column Division) at Metrohm International Headquarters, Herisau, Switzerland.

Best practice for electrodes in Karl Fischer titration

Best practice for electrodes in Karl Fischer titration

Have you ever asked yourself why you need an electrode for the endpoint detection in Karl Fischer (KF) titration? Theoretically, the endpoint of a Karl Fischer titration could be determined based on the color change of the reagent. However, if accuracy and reproducibility are important, endpoint detection with a double Pt electrode is a much better choice.

As the indicator electrode detects the endpoint, you can imagine that the results depend highly on the condition of the electrode. In coulometry, an additional electrode (generator electrode) is used to generate the iodine needed for the titration. Both electrode types (i.e. indicator and generator electrode) need to be kept in good shape to guarantee the correct results. It goes without saying that cleaning, storage, maintenance, and checks of the KF electrodes are important factors for success. This blog post takes a closer look at these topics.

Did you catch our series about frequently asked questions in Karl Fischer titration? Find them here!

Cleaning

Indicator electrode

Double  Pt-wire or double Pt-ring electrodes can be easily cleaned with an abrasive cleaning agent like aluminum oxide powder or toothpaste. After cleaning, rinse the electrode well with water and let it dry before mounting it in a titration cell. Check out our video below for more tips and tricks about the proper cleaning procedure for Karl Fischer titration indicator electrodes.

Take special care not to bend the Pt pins of the double Pt-wire electrode. Bending the pins can lead to tiny cracks in the glass body of the electrode. Over time, reagent can flow into the electrode and lead to corrosion (short circuit). If this happens, the electrode is beyond repair and needs replacement. Alternatively, a double Pt-ring electrode can be used instead. Problems with bent pins are then a thing of the past.
Generator electrode
Without diaphragm
Rinse generator electrodes without diaphragms with water, or if the contaminant is not water soluble, then rinse with a suitable organic solvent. If the anode or the cathode of the generator electrode shows discoloration or deposits that cannot be removed with rinsing, the electrode can then be cleaned with concentrated nitric acid (65%). Be aware that nitric acid is a strong oxidizing agent and must be handled carefully according to relevant safety regulations and instructions. Remember to first mount the green protection cap on the connector to avoid corrosion caused by fumes of nitric acid. Afterwards, rinse the electrode with water and finally with methanol.
With diaphragm
To remove salt-like residues, the generator electrode with diaphragm can be rinsed with water. Oily contamination can be rinsed off with an organic solvent (e.g. hexane). Sticky residues on the diaphragm can be removed in the following way: 

  1. Mount the green protection cap on the connector of the electrode.
  2. Place the electrode in an upright position (e.g. in an Erlenmeyer flask) and add a few milliliters of concentrated nitric acid (65%) in the cathode chamber. Let the acid flow through the diaphragm.
  3. Fill the cathode chamber with water and let it flow through the diaphragm to remove the nitric acid. Repeat this step two or three times. A simple way to see whether another rinsing step is required is by performing a quick check of the pH value at the cathode using pH indication paper.
  4. Finally, fill the cathode chamber with methanol and let it flow out.

Now the generator electrode is as good as new and ready for use in a titration cell again.

Maintenance

Except for the generator electrode with diaphragm, KF electrodes are maintenance free. However, the catholyte filled in the generator electrode with diaphragm can decompose over time. To avoid any influence of the decomposition products on the results, exchange the catholyte on a regular basis according to the manufacturer’s recommendations.

Storage

Unlike pH electrodes, KF electrodes do not contain a glass membrane that could potentially dry out. Therefore, no special solution is required in which to store KF electrodes. If you use the electrodes frequently, it is recommended to keep the electrodes mounted in the titration cell and immersed in the KF reagent. Alternatively, all KF electrodes (indicator and generator electrodes) can be stored dry.

What to check for

It is recommended to check the complete titration setup instead of only the electrode(s).

Volumetry

Carry out a threefold titer determination using either a liquid or a solid water standard suitable for volumetry and calculate the mean value of the titer. Then, determine the water content of a water standard (also via triplicate determination). Make sure that you do not use the same standard as for the titer determination but use a different batch of the standard or even a completely different standard. Calculate the water content and compare it to the certified water content of the standard.

If the recovery is determined to be in the range of 97–103%, the titration system (including the electrode) is working fine. Finding values outside this range means that there is something wrong with the titration system or with the determination procedure. Results of the sample analysis would very likely also deviate from the real water content. Therefore, it is important to find the reason for values that are too high or too low. Sometimes the reason for deviations is just an air bubble in the dosing cylinder or due to an exhausted molecular sieve. However, if you do not find the reason, do not hesitate to contact your local Metrohm agency.

Coulometry

Water standards with lower water contents (0.1%) are available to properly check the health of coulometric titration systems. Carry out a water content determination in triplicate with such a standard. Calculate the recovery with the obtained results and the certified water content of the standard.

A recovery value in the range between 97–103% means that everything is fine with the system and that the electrodes work as expected. As with volumetry, in coulometry it is important to find the reason for any deviating recovery values. Make sure that you find and eliminate the problem to obtain correct results for your samples.

What you should avoid

  • Do not use solvents that contain ketones or aldehydes (e.g. denatured ethanol) to clean KF electrodes or any KF accessories.
  • Do not treat KF electrodes in an ultrasonic bath. This might destroy the electrode.
  • For drying, use a maximum temperature of 50 °C. Higher temperatures might damage the electrode.
  • Do not bend the Pt pins of the double Pt-wire electrode.

Summary

As you can see, keeping your KF electrodes in good shape is actually very simple. Regular cleaning helps to avoid erroneous results and ensures that your Karl Fischer electrodes will work for a long time.

Best practice for electrodes in titration

Treat your sensors right!
Post written by Michael Margreth, Sr. Product Specialist Titration (Karl Fischer Titration) at Metrohm International Headquarters, Herisau, Switzerland.
Best practice for separation columns in ion chromatography (IC) – Part 3

Best practice for separation columns in ion chromatography (IC) – Part 1

The high performance ion chromatography (IC) separation column is often referred to as the «heart» of the ion chromatograph. The reason for this denomination is straightforward: the column is responsible for the separation of the analytes of interest from each other as well as from interfering sample matrix ions. The unique separation capabilities of IC columns allow the determination of multiple analytes within one run. In this blog series, we will share what is required to ensure the proper operation of an IC column and how to maximize the column lifetime.

Standard operating conditions

To begin with, the standard operating conditions should be considered, such as the eluent (mobile phase) composition, the eluent flow rate, the column oven temperature, and the detection method. These standard conditions are specific for each individual column type and correspond to the conditions that work best with the application the column is intended for. Every analytical column sold by Metrohm is delivered with a certificate of analysis (CoA) which is recorded under standard operating conditions. In the following table, you can find the standard operating conditions for three different Metrohm IC columns.

Column Metrosep C 6 – 150/4.0 Metrosep A Supp 17 – 150/4.0 Metrosep Carb 2 – 150/4.0
Eluent composition 1.7 mmol/L nitric acid

1.7 mmol/L dipicolinic acid

5.0 mmol/L sodium carbonate

0.2 mmol/L sodium bicarbonate

100 mmol/L sodium hydroxide

10 mmol/L sodium acetate

Flow rate 0.9 mL/min 0.6 mL/min 0.5 mL/min
Oven Temperature 30 °C 25 °C 30 °C
Detection Non-suppressed conductivity Suppressed conductivity Amperometric detection
Analytes Lithium, sodium, ammonium, potassium, calcium, magnesium Fluoride, chloride, nitrite, bromide, nitrate, sulfate, phosphate Inositol, arabitol, sorbitol, glucose, xylose, fructose, lactose, sucrose

Equilibration

Next to the standard operating conditions, the start-up parameters play an important role for the lifetime of the separation column. High mechanical and thermal stress is a frequent reason for decreased column lifetime. It is therefore recommended to slowly increase the eluent flow rate to the column and to avoid thermal and pressure shocks. For more information regarding specifics, please refer to the recommended equilibration conditions in the corresponding column leaflet.

Operating limits

Based on the different stationary phase designs that are required to achieve different selectivities and guarantee a wide range of IC applications, the chemical and physical properties of the ion exchangers can vary. Therefore, different operating limits are recommended for various column types and set according to stress tests performed during the development of a product. Before beginning to optimize an application, please refer to the respective column leaflet for the corresponding operating limits to achieve optimal results and guarantee a long column lifetime.

The chemistry of the ion exchanger defines the limits of operating temperature, eluent pH, and organic modifiers that can be present in the eluent. These values are valid for every dimension of a column type (e.g., the Metrosep C 6). Exceeding these limits can strongly affect column performance and, in the worst case, lead to irreversible damage. Tolerated flow rates and maximum pressure correlate with physical properties and of course column dimension—therefore these limits are set for every column dimension (e.g., Metrosep C 6 – 100/4.0 vs. Metrosep C 6 – 250/2.0).

Are you searching for a specific IC column for your research? Then check out the Metrohm Column Finder!

Information leaflet

In addition to the CoA, a lot of necessary valuable information can be found in the respective column leaflet. Instructions about equilibration, regeneration, operating limits, and much more can be found in the column leaflet which is provided for every IC column type offered by Metrohm.

You can download an example of a column leaflet here, where the information is provided in several different languages (DE, EN, FR, and ES).

Storage

Depending on the operation conditions and the properties of the ion exchanger material, different storage conditions are recommended for different column types. These conditions (storage temperature and storage eluent) are described on the respective column leaflet and should be strictly followed.

Please note that storing a new column for long time periods as well as frequently switching between storage and operation can actually be more stressful to the column than operating it 24/7! It is therefore not recommended to stock a column for a long time without using it.

Column guards

To protect your separation column from sample contamination and to extend its lifetime, it is critical to use a guard column. The guard column should be exchanged at regular intervals—a general rule of thumb is that about four guard columns will be used over the lifetime of an IC column.

Guard columns are offered in two designs: «on column» (e.g. Metrosep A Supp 17 Guard/4.0, on the left) and as a separate guard that is connected to the IC column with a capillary (e.g. Metrosep A Supp 17 S-Guard/4.0, on the right). For microbore separation columns, the corresponding microbore column guards are recommended (e.g. Metrosep A Supp 10 Guard/2.0).

By default, a guard column with the same material as the column should be used. However, for special applications, combining different ion exchange materials by using a different guard column can help to optimize the separation. One such example is in the case of sulfate and sulfite, as shown below.

Pulsation absorber

With the exception of the Eco IC product line, all Metrohm ion chromatography instruments are equipped with a pulsation absorber. For the Eco IC, it is strongly recommended to add the pulsation absorber as an option. As mentioned earlier, IC columns do not like repeated mechanical stress, especially those based on polyvinyl alcohol or polymethacrylate stationary phases. Therefore, the pulsation absorber is a useful tool to protect the column from possible pressure fluctuations in the system and enhance the column lifetime.

Find out more about the full line of Metrohm ion chromatography products and accessories on our website!

Chemical quality

In ion chromatography, the ratio between the volume of sample and the volume of eluent that flows through the column is very small, usually in the range of 1:1000. Therefore the quality of the chemical reagents used in the eluent preparation plays a crucial role for the column lifetime. To guarantee optimal performance, use chemicals denominated as «for IC» whenever possible as they are tested particularly on impurities (e.g. metals) that could harm the chromatographic column.

For dilution of the eluent components, ultrapure water is typically used in ion chromatography. To ensure good chromatographic results, the ultrapure water should have a specific resistance greater than 18 MΩ · cm and be free from particles. The ultrapure water is filtered through a 0.45 µm filter and treated with UV. Modern ultrapure water sources for laboratory use guarantee this level of water quality (Type I).

Sample preparation

With the Metrohm Inline Sample Preparation (MISP) options, Metrohm provides a large number of sample preparation techniques that are beneficial to the separation column, as well as analysts. Instead of loading the full sample onto the column, these techniques ensure the reduction of several sample matrix effects, thus avoiding potential harm to the column.

One of the most prominent MISP techniques is Inline Ultrafiltration (illustrated here), which efficiently removes particles from the sample in a fully automated way, before they ever reach the column. In that way, column blockage from dirty samples can be avoided without any manual effort.

iColumn features

All Metrosep columns offered by Metrohm are equipped with an intelligent chip containing useful information about various column operation conditions (e.g., equilibration, standard operation conditions, operation limits, and more) and tracks certain parameters (e.g., set to work, number of injections, number of working hours, and maximum working values such as pressure and flow rate) over the column lifetime. It is beneficial to attach the column chip to the chip reader as illustrated here for proper monitoring as well as support.

Column shelf life

How many injections are possible on a specific column? Unfortunately, it is not possible to deliver an exact answer to this question. This is because the column lifetime strongly depends on the sample matrix and elution conditions. Due to the multitude of different applications and samples that can be covered with a single IC column, it is not possible to predict the column life for every application and sample type.

During column development, several endurance tests are performed under standard conditions with a guard column using appropriate standards. Under these conditions, the column must withstand at least 2000 injections.

The History of Metrohm IC

Metrohm ion chromatography: bringing top quality and exceptional analytical performance to the lab since 1987. 

Post written by Dr. Vincent Diederich (Product Manager IC Columns) and Dr. Anne Katharina Riess (Head of Column Division) at Metrohm International Headquarters, Herisau, Switzerland.

Analysis of prebiotics with IC-PAD: Improving AOAC 2001.02

Analysis of prebiotics with IC-PAD: Improving AOAC 2001.02

Our diet is critical for our health. In the past several years, interest has increased in food additives and dietary supplements such as prebiotics like β-galactooligosaccharides (GOSs). The determination of total GOS contents in food and supplements is essential to fulfill strict food labeling and safety requirements. The most widely used method for total GOS determination is based on enzymatic hydrolysis to break down the complex molecules into simple carbohydrates prior to their chromatographic analysis. This article outlines the advantage of using an improvement to AOAC Method 2001.02 using ion chromatography with amperometric detection (IC-PAD) and full sample automation after enzymatic hydrolysis.

What are GOSs?

GOSs are chains of galactose units with an optional glucose end. They are often naturally present in small amounts in various foods and beverages.

Initially discovered as major constituents of human breast milk (present up to 12 g/L), GOSs are added as a prebiotic supplement to infant formulas. They show bifidogenic effects, meaning they support growth and well-being of non-pathogenic gut bacteria.

GOS supplements are available either raw, or as concentrated powders or syrups, and are subsequently used by food manufacturers to enrich consumer products or sold as supplements.

GOS labeling requirements

The ongoing growth of global prebiotic and GOS markets is a result of increasing consumer awareness regarding healthy eating. Similarly, increased demand regarding food quality has led to stricter, more comprehensive rules for food labeling and safety (e.g., EU 1169/2011 and  EU 2015/2283). The determination of total GOS contents in food, supplements, or raw products is thus essential to fulfill such requirements.

Studies about GOS health effects recommend maximum doses under 30 g per day, though this is much stricter for infant formulas. Otherwise, there are no other limits regarding GOS content in food or as nutritional supplements.

AOAC 2001.02

The most widely used method to measure total GOSs in food products is the standard method AOAC 2001.02. This method is based on the extraction of GOS from a sample followed by enzymatic hydrolysis of the oligosaccharides into monosaccharides and their subsequent analyses with high performance anion exchange chromatography with pulsed amperometric detection.

Figure 1. Schematic for determination of total GOS contents using ion chromatography with pulsed amperometric detection (IC-PAD) according to AOAC 2001.02, and an optimized method from Metrohm (in green). Chromatography for anions in AOAC is referred as HPAEC (high performance anion exchange chromatography) but is simplified here to the generic term of IC.

In AOAC, chromatography for anions is referred to as HPAEC (high performance anion exchange chromatography) but here we will simplify this to the generic term of IC.

The key to AOAC 2001.02 is the comparison of a control solution with one which has been treated and hydrolyzed with an enzyme (β-galactosidase). The enzyme catalyzes the splitting of glycosidic bonds and hydrolyzes GOSs and lactose into glucose and galactose. The concentration differences of free galactose and lactose determined in these two solutions is used to calculate the total GOSs (Figure 1).

Improvements to the AOAC Method

The sample preparation for AOAC 2001.02 is rather complex: one shortcoming is the incubation of the reference solution with the deactivated enzyme (which is rather expensive) to determine the initial carbohydrate concentrations (Figure 1) rather than using the pure extract. Another critical point is the sample dilution procedure, which is supposed to be done in acetonitrile, while standards are based on ultrapure water.

Here, the focus was to simplify the entire procedure to increase the ease of use and the overall efficiency of the method.

The improved method for total GOS content analysis uses the extract for measuring of the initial glucose, galactose, and lactose concentrations (Figure 1 Assay 1). However, the deactivated enzyme was not used, and instead comparisons were made to see if its presence had any effect on the results. This step was eliminated after proving results equivalent to AOAC 2001.02 Assay 1 (with the deactivated enzyme), but chemical expenses and additional manual work are reduced. The total GOS content is therefore calculated from the analyte concentrations in Assay 1 (without any enzyme) and Assay 2 (extract with the active enzyme) (Figure 2).

Figure 2. Overlaid chromatograms of Bimuno (prebiotic supplement), untreated (black) and treated with enzyme (orange).

Want to know more details about the application? Download our free Application Note AN-P-087 about total GOS analysis in foods with ion chromatography!

Aside from the enzyme usage, the official AOAC method for analysis of total GOSs suggests that standards be prepared in ultrapure water (UPW) while samples are to be diluted with 20% acetonitrile. A control experiment was performed to compare results between:

  • Dilutions in UPW evaluated with UPW calibration (“UPW option”)
  • Dilutions in acetonitrile evaluated with UPW calibration (AOAC 2001.02)
  • Dilutions in acetonitrile evaluated with acetonitrile calibration (“ACN option”)

Reproducibility of total GOS contents was compared among the three options, with the UPW and AOAC preparation options exhibiting similar results. The ACN option resulted in lower total GOS contents than the others. Additionally, the acetonitrile did not seem to lend a stabilizing effect to the samples. This supports the improvement of the AOAC method by performing sample dilutions with UPW instead of acetonitrile, saving unnecessary reagents and limiting the chemical imprint of the analysis.

Results

Overall, the satisfying variability, target and spike recoveries (Application Note AN-P-087), together with the interference tests proved the modified method as valuable and robust. With limits of detection (LODs) of 0.1 mg/L (galactose) and 0.2 mg/L (glucose, lactose) in solution, even low total GOS contents can be determined with high precision.

Summary

As a multicomponent method, ion chromatography with amperometric detection is a very selective, sensitive, and robust analysis method for carbohydrates without any additional derivatization steps. In combination with enzymatic treatment, even more complex carbohydrates can be quantified.

This research presents an update to the standard AOAC method for total GOS determination in foodstuffs. With the same principle (enzymatic hydrolysis of complex GOS molecules followed by chromatographic analysis of simple carbohydrates), analytical method efficiency was improved in favor of laboratory time and running costs. Additional automation steps (e.g., Metrohm Inline Dilution and automatic calibrations) can further improve the method efficiency.

Want more information about the simplified method for total GOSs via IC-PAD? More details about the improvement of AOAC method 2001.02 by reducing manual laboratory work and eliminating expensive reagents can be found in our article published in The Column from LC/GC (2021): Improving on AOAC 2001.02: GOS Determination in Foods Using HPAEC–PAD.

Read our article in LC/GC The Column (2021)

Improving on AOAC 2001.02: GOS Determination in Foods Using HPAEC–PAD

Post written by Dr. Alyson Lanciki, Scientific Editor at Metrohm International Headquarters, Herisau, Switzerland.

«Analyze This»: 2020 in review

«Analyze This»: 2020 in review

I wanted to end 2020 by thanking all of you for making «Analyze This» – the Metrohm blog for chemists such a success! For our 60th blog post, I’d like to look back and focus on the wealth of interesting topics we have published this year. There is truly something for everyone: it doesn’t matter whether your lab focuses on titration or spectroscopic techniques, or analyzes water samples or illicit substances – we’ve got you covered! If you’re looking to answer your most burning chemical analysis questions, we have FAQs and other series full of advice from the experts. Or if you’re just in the mood to learn something new in a few minutes, there are several posts about the chemical world to discover.

We love to hear back from you as well. Leaving comments on your favorite blog posts or contacting us through social media are great ways to voice your opinion—we at Metrohm are here for you!

Finally, I wish you and your families a safe, restful holiday season. «Analyze This» will return on January 11, 2021, so subscribe if you haven’t already done so, and bookmark this page for an overview of all of our articles grouped by topic!

Stay healthy, and stay curious.

Best wishes,

Dr. Alyson Lanciki, Scientific Editor, Metrohm AG

Quickly jump directly to any section by clicking a topic:

Customer Stories

We are curious by nature, and enjoy hearing about the variety of projects where our products are being used! For some examples of interesting situations where Metrohm analytical equipment is utilized, read on.

From underwater archaeological research to orbiting Earth on the International Space Station, Metrohm is there! We assist on all types of projects, like brewing top quality beers and even growing antibiotic-free shrimp – right here in Switzerland.

Interested in being featured? Contact your local Metrohm dealer for details!

Titration

Metrohm is the global market leader in analytical instruments for titration. Who else is better then to advise you in this area? Our experts are eager to share their knowledge with you, and show this with the abundance of topics they have contributed this year to our blog.

For more in-depth information about obtaining the most accurate pH measurements, take a look at our FAQ about pH calibration or read about avoiding the most common mistakes in pH measurement. You may pick up a few tips!

Choose the best electrode for your needs and keep it in top condition with our best practices, and then learn how to standardize titrant properly. Better understand what to consider during back-titration, check out thermometric titration and its advantages and applications, or read about the most common challenges and how to overcome them when carrying out complexometric titrations

If you are interested in improving your conductivity measurements, measuring dissolved oxygen, or the determination of oxidation in edible fats and oils, check out these blog posts and download our free Application Notes and White Papers!

Finally, this article about comprehensive water analysis with a combination of titration and ion chromatography explains the many benefits for laboratories with large sample loads. The history behind the TitrIC analysis system used for these studies can be found in a separate blog post.

Karl Fischer Titration

Metrohm and Karl Fischer titration: a long history of success. Looking back on more than half a century of experience in KFT, Metrohm has shaped what coulometric and volumetric water analysis are today.

Aside from the other titration blog posts, our experts have also written a 2-part series including 20 of the most frequently asked questions for KFT arranged into three categories: instrument preparation and handling, titration troubleshooting, and the oven technique. Our article about how to properly standardize Karl Fischer titrant will take you step by step through the process to obtain correct results.

For more specific questions, read about the oven method for sample preparation, or which is the best technique to choose when measuring moisture in certain situations: Karl Fischer titration, near-infrared spectroscopy, or both?

Ion Chromatography (IC)

Ion chromatography has been a part of the Metrohm portfolio since the late 1980s. From routine IC analysis to research and development, and from stand-alone analyzers to fully automated systems, Metrohm has provided IC solutions for all situations. If you’re curious about the backstory of R&D, check out the ongoing series about the history of IC at Metrohm.

Metrohm IC user sitting at a laboratory bench.

Common questions for users are answered in blog posts about IC column tips and tricks and Metrohm inline ultrafiltration. Clear calculations showing how to increase productivity and profitability in environmental analysis with IC perfectly complement our article about comprehensive water analysis using IC and titration together for faster sample throughput.

On the topic of foods and beverages, you can find out how to determine total sulfite faster and easier than ever, measure herbicides in drinking water, or even learn how Metrohm IC is used in Switzerland to grow shrimp!

Near-Infrared Spectroscopy (NIRS)

Metrohm NIRS analyzers for the lab and for process analysis enable you to perform routine analysis quickly and with confidence – without requiring sample preparation or additional reagents and yielding results in less than a minute. Combining visible (Vis) and near-infrared (NIR) spectroscopy, these analyzers are capable of performing qualitative analysis of various materials and quantitative analysis of a number of physical and chemical parameters in one run.

Our experts have written all about the benefits of NIR spectroscopy in a 4-part series, which includes an explanation of the advantages of NIRS over conventional wet chemical analysis methods, differences between NIR and IR spectroscopy, how to implement NIRS in your laboratory workflow, and examples of how pre-calibrations make implementation even quicker.

A comparison between NIRS and the Karl Fischer titration method for moisture analysis is made in a dedicated article.

A 2-part FAQ about NIRS has also been written in a collaboration between our laboratory and process analysis colleagues, covering all kinds of questions related to both worlds.

Raman Spectroscopy

This latest addition to the Metrohm family expands the Metrohm portfolio to include novel, portable instruments for materials identification and verification. We offer both Metrohm Raman as well as B&W Tek products to cover a variety of needs and requirements.

Here you can find out some of the history of Raman spectroscopy including the origin story behind Mira, the handheld Raman instrument from Metrohm Raman. For a real-world situation involving methamphetamine identification by law enforcement and first responders, read about Mira DS in action – detecting drugs safely in the field.

Mira - handheld Raman keeping you safe in hazardous situations.

Are you looking for an easier way to detect food fraud? Our article about Misa describes its detection capabilities and provides several free Application Notes for download.

Process Analytics

We cater to both: the laboratory and the production floor. The techniques and methods for laboratory analysis are also available for automated in-process analysis with the Metrohm Process Analytics brand of industrial process analyzers.

Learn about how Metrohm became pioneers in the process world—developing the world’s first online wet chemistry process analyzer, and find out how Metrohm’s modular IC expertise has been used to push the limits in the industrial process optimization.

Additionally, a 2-part FAQ has been written about near-infrared spectroscopy by both laboratory and process analysis experts, which is helpful when starting out or even if you’re an advanced user.

Finally, we offer a 3-part series about the advantages of process analytical technology (PAT) covering the topics of process automation advantages, digital networking of production plants, and error and risk minimization in process analysis.

Voltammetry (VA)

Voltammetry is an electrochemical method for the determination of trace and ultratrace concentrations of heavy metals and other electrochemically active substances. Both benchtop and portable options are available with a variety of electrodes to choose from, allowing analysis in any situation.

A 5-part series about solid-state electrodes covers a range of new sensors suitable for the determination of «heavy metals» using voltammetric methods. This series offers information and example applications for the Bi drop electrode, scTrace Gold electrode (as well as a modified version), screen-printed electrodes, and the glassy carbon rotating disc electrode.

Come underwater with Metrohm and Hublot in our blog post as they try to find the missing pieces of the ancient Antikythera Mechanism in Greece with voltammetry.

If you’d like to learn about the combination of voltammetry with ion chromatography and the expanded application capabilities, take a look at our article about combined analysis techniques.

Electrochemistry (EC)

Electrochemistry plays an important role in groundbreaking technologies such as battery research, fuel cells, and photovoltaics. Metrohm’s electrochemistry portfolio covers everything from potentiostats/galvanostats to accessories and software.

Our two subsidiaries specializing in electrochemistry, Metrohm Autolab (Utrecht, Netherlands) and Metrohm DropSens (Asturias, Spain) develop and produce a comprehensive portfolio of electrochemistry equipment.

This year, the COVID-19 pandemic has been at the top of the news, and with it came the discussion of testing – how reliable or accurate was the data? In our blog post about virus detection with screen-printed electrodes, we explain the differences between different testing methods and their drawbacks, the many benefits of electrochemical testing methods, and provide a free informative White Paper for interested laboratories involved in this research.

Our electrochemistry instruments have also gone to the International Space Station as part of a research project to more efficiently recycle water on board spacecraft for long-term missions.

The History of…

Stories inspire people, illuminating the origins of theories, concepts, and technologies that we may have become to take for granted. Metrohm aims to inspire chemists—young and old—to be the best and never stop learning. Here, you can find our blog posts that tell the stories behind the scenes, including the Metrohm founder Bertold Suhner.

Bertold Suhner, founder of Metrohm.

For more history behind the research and development behind Metrohm products, take a look at our series about the history of IC at Metrohm, or read about how Mira became mobile. If you are more interested in process analysis, then check out the story about the world’s first process analyzer, built by Metrohm Process Analytics.

Need something lighter? Then the 4-part history of chemistry series may be just what you’re looking for.

Specialty Topics

Some articles do not fit neatly into the same groups as the rest, but are nonetheless filled with informative content! Here you can find an overview of Metrohm’s free webinars, grouped by measurement technique.

If you work in a regulated industry such as pharmaceutical manufacturing or food and beverage production, don’t miss our introduction to Analytical Instrument Qualification and what it can mean for consumer safety!

Industry-focused

Finally, if you are more interested in reading articles related to the industry you work in, here are some compilations of our blog posts in various areas including pharmaceutical, illicit substances, food and beverages, and of course water analysis. More applications and information can be found on our website.

Food and beverages
All of these products can be measured for total sulfite content.

Oxidation stability is an estimate of how quickly a fat or oil will become rancid. It is a standard parameter of quality control in the production of oils and fats in the food industry or for the incoming goods inspection in processing facilities. To learn more about how to determine if your edible oils are rancid, read our blog post.

Determining total sulfite in foods and beverages has never been faster or easier than with our IC method. Read on about how to perform this notoriously frustrating analysis and get more details in our free LC/GC The Column article available for download within.

Measuring the true sodium content in foodstuff directly and inexpensively is possible using thermometric titration, which is discussed in more detail here. To find out the best way to determine moisture content in foods, our experts have written a blog post about the differences between Karl Fischer titration and near-infrared spectroscopy methods.

To determine if foods, beverages, spices, and more are adulterated, you no longer have to wait for the lab. With Misa, it is possible to measure a variety of illicit substances in complex matrices within minutes, even on the go.

All of these products can be measured for total sulfite content.

Making high quality products is a subject we are passionate about. This article discusses improving beer brewing practices and focuses on the tailor-made system built for Feldschlösschen, Switzerland’s largest brewer.

Pharmaceutical / healthcare

Like the food sector, pharmaceutical manufacturing is a very tightly regulated industry. Consumer health is on the line if quality drops.

Ensuring that the analytical instruments used in the production processes are professionally qualified is a must, especially when auditors come knocking. Find out more about this step in our blog post about Analytical Instrument Qualification (AIQ).

Moisture content in the excipients, active ingredients, and in the final product is imperative to measure. This can be accomplished with different analytical methods, which we compare and contrast for you here.

The topic of virus detection has been on the minds of everyone this year. In this blog post, we discuss virus detection based on screen-printed electrodes, which are a more cost-effective and customizable option compared to other conventional techniques.

Water analysis

Water is our business. From trace analysis up to high concentration determinations, Metrohm has you covered with a variety of analytical measurement techniques and methods developed by the experts.

Learn how to increase productivity and profitability in environmental analysis laboratories with IC with a real life example and cost calculations, or read about how one of our customers in Switzerland uses automated Metrohm IC to monitor the water quality in shrimp breeding pools.

If heavy metal analysis is what you are interested in, then you may find our 5-part series about trace analysis with solid-state electrodes very handy.

Unwanted substances may find their way into our water supply through agricultural practices. Find out an easier way to determine herbicides in drinking water here!

Water is arguably one of the most important ingredients in the brewing process. Determination of major anions and cations along with other parameters such as alkalinity are described in our blog post celebrating International Beer Day.

All of these products can be measured for total sulfite content.
Illicit / harmful substances

When you are unsure if your expensive spices are real or just a colored powder, if your dairy products have been adulterated with melamine, or fruits and vegetables were sprayed with illegal pesticides, it’s time to test for food fraud. Read our blog post about simple, fast determination of illicit substances in foods and beverages for more information.

Detection of drugs, explosives, and other illegal substances can be performed safely by law enforcement officers and first responders without the need for a lab or chemicals with Mira DS. Here you can read about a real life training to identify a methamphetamine laboratory.

Drinking water regulations are put in place by authorities out of concern for our health. Herbicides are important to measure in our drinking water as they have been found to be carcinogenic in many instances.

Post written by Dr. Alyson Lanciki, Scientific Editor at Metrohm International Headquarters, Herisau, Switzerland.